Voltage (V) |
Ranges (Auto
or Manual)(6 range) |
15V 30V 60V 150V 300V
600V RMS |
Display Resolution |
4-1/2 digits |
Scaling Factors |
0.1 to 1000.0 |
Operating Principle |
True RMS or mean Value rectification |
Crest Factor |
Up to 3 |
Frequency Range |
DC and 10Hz to 100kHz (300kHz
Only CP-620-2) |
Accuracy(23℃±5℃) |
47 to 63Hz |
±0.05% reading ±0.05%
range ±10mV |
DC and 10Hz to 100k Hz |
±0.2% reading ±0.1% range
±0.1% / kHz |
|
50kHz to 300kHz
( Only CP-620-2 )
(建议外加高频电流感测器) |
±0.3% reading ±0.2% range ±0.05%
/kHz |
Maximum Input |
Continuous |
600Vrms / 1800Vpk |
Effect of Common
Mode Voltage |
600Vrms at 60Hz |
Less than ± 0.05% of range |
|
Input impedance |
1Mohm and 10pF on all
ranges |
Current (A) |
Ranges (Auto or
Manual)
(7 ranges 1-2-5 sequences) |
CP-620-2 CP-660-2 |
0.2A 0.5A 1A 2A 5A 10A 20A |
CP-660-5 |
0.5A 1A 2A 5A 10A 20A 50A |
EXT IN (7
Range) (RMS) |
20mV 50mV 100mV 200mV
500mV 1V 2V |
Display Resolution |
4-1/2 digits |
Scaling Factor |
0.01 to 1000.0 |
Operating Principle |
True RMS |
Frequency
Range |
DC and 10Hz to 100kHz(300kHz
Only CP-620-2) |
Crest Factor |
Up to 3 |
Accuracy(23℃±5℃) |
47 to 63Hz |
±0.05% reading ± 0.05%range
± 0.1mA |
DC and 10Hz to 100kHz |
±0.2% reading ± 0.1% range
± 0.1%/kHz |
|
50kHz to 300kHz
( Only CP-620-2 )
(建议外加高频电流感测器) |
±0.3% reading ± 0.2% range ±
0.1% /kHz |
Maximum Input |
Continuous |
20Arms / 50Arms (60Apk
/ 150Apk) |
Effects of Common
Mode Voltage |
600Vrms at 60Hz |
Less than ± 0.05% of range |
- |
Input impedance |
CP-660-2 |
10mohm on all ranges(CP-620
not) |
CP-660-5 |
3.5mohm on all ranges |
Power (W) |
Ranges (Auto or
Manual)
(42 ranges corresponding to V and A) |
3W to 30kW (with scaling to
99.999MW) |
Display Resolution |
4-1/2 digits |
Operating principle |
Analog multiplier Circuit |
Frequency
Range |
DC and 10Hz to 100kHz(300kHz
Only CP-620) |
Crest Factor |
Up to 3 |
Accuracy (23℃±5℃) |
47Hz to 63Hz |
±0.1% reading ± 0.1% range
± 5mW |
DC and 10Hz to 50kHz |
±0.2% reading ± 0.1% range
± 0.1%/kHz |
Total Harmonic
Distortion (THD)
(CP-660-2 CP-660-5 Only) |
Ranges (Auto) |
10%. 100% |
Display resolution |
4 digits (0.001% ~ 99.99%) |
Frequency Range , Fundamental
(Automatic Tuned) |
45Hz ~ 65Hz. 370Hz ~ 430Hz |
Crest Factors |
Up to 3 |
Fundamental Rejection |
80db and above |
Residual Distortion and noise |
Less than 0.1% (Sine Wave) |
Detection Type (THD-R)
(THD-F=THD-R/√(1-(THD-R)2)
( 频宽设定为5kHz ) (F 为基本波) 详细资料请参考技术支援 |
True RMS
|
Low pass filter band wide (3
Range) |
500Hz , 5kHz , pass(80kHz) |
Measuring Input |
Voltage or Current |
Frequency
(Hz) |
Ranges (Auto 4 Range) |
100Hz. 1kHz. 10kHz. 100kHz |
Display Resolution / Maximum
resolution |
5 digits / 0.001Hz |
Operating Principle |
Reciprocal Counting method |
Accuracy (23±5℃) |
±0.01% ±1C |
Measuring Input |
Voltage or Current |
Peak Current
(+pk)
(-pk)
(P-P) |
Ranges (Auto or Manual)
(7 Range) |
CP-600-2 CP-660-2 |
±0.6A 1.5A 3A 6A 15A 30A 60A (CP-620-2) |
CP-600-5 CP-660-5 |
±1.5A 3A 6A 15A 30A 60A 150A |
Display Resolution |
4-1/2 digits |
Frequency Range |
10Hz to 100kHz |
Accuracy |
±1% reading ±1%range ±1%/kHz |
Crest Factor
(CF) |
Ranges |
1.000 to 9.999 |
Frequency Range |
10Hz to 100kHz |
Accuracy |
±1%reading ±1%range ±1%/kHz |
Measuring Input |
Voltage or Current |
Inrush Current
(Ins) |
Ranges (7 range / Int shunt) |
60mA to 60Apk(20A)/150mA to 150A peak(50A) |
Accuracy |
±3% of selected range |
Integrator
(INTG) |
Function |
Wh. VAh. VArh. Ah |
Display Resolution |
6 digits (Max 999.999M) |
Unit |
m.k.M |
Integration Interval |
0.3 second |
Elapsed Time Display |
5 digits 1 minute to 999 hours 59
minute |
Timer accuracy |
±0.01% |
Interfaces
(Optional) |
GPIB |
Specifications |
Standard IEEE 488.2 |
Function |
SH1 AH1 T6 L4 SR1 PL1 PP0 DC1 DT1
C0 |
RS-232 |
Mode |
Start stop synchronization |
Baud rate |
1200. 2400. 4800. 9600bps |
Print Port |
Specifications |
Standard Centronics (IBM-PC
Compatible) |
Driver |
EPSON LQ Series and Compatible |
Print Interval |
1 second to 24 Hours(1 second STEP) |
Print Count |
1~9999 |
Print Format |
Panel display or All Parameter |