- Current Range :
- 100 mV/A : 100 mA to 10 A peak
- 10 mV/A : 1 to 100 A peak
- Output Signal :
- mV output (1000 mV peak max)
- AC Current Accuracy :
- 100 mV/A (50 mA to 10 A peak) : 3% of reading ¡Ó
50mA
- 10 mV/A (500 mA to 40 A peak) : 4% of reading ¡Ó
50mA
- 10 mV/A (40 mA to 100 A peak) : 15% max. at 100 A
- Phase Shift* :
- < 1゜from DC to 65 Hz on 10 mV/A
- < 1.5゜from DC to 65 Hz on 100 mV/A
- Frequency Range: DC to 100kHz (-3 dB with current derating)
- Noise: Range 10 mV/A: 480 µV
- Range 100 mV/A: 3 mV
- Slew Rate: Range 10 mV/A: 20mV/µs
- Range 100 mV/A: 0.3 V/µs
- Load Impedance :
- Insertion Impedance (50/60 Hz) :
- 100 mV/A:0.01£[
- 10 mV/A: 0.01£[
|
- Rise or Fall Time :
- Range 100 mV/A : 3µs
- Range 10 mV/A : 4µs
- Working Voltage : 600 Vrms max
- Common Mode Voltage : 600 Vrms max
- Influence of Adjacent Conductor : < 0.2 mA/A AC
- Influence of Conductor Position in Jaw : 0.5% reading at 1 kHz
- Battery :
- 9 V alkaline (NEDA 1604A, IEC 6LR61)
- Low Battery :
- Overload Indication :
- Red LED indicates input greater than the selected range
- Typical Consumption : 8.6 mA
- Battery Life : 55 hours typical
(*Reference Conditions: 23˚C กำ 5˚K, 20 to 75%RH, DC to 1 kHz,
probe zeroed,
1 minute warm- up, battery at 9 V ¡Ó 0.1 V, external magnetic
field < 40 A/m,
no DC component, no external current carrying conductor, 1 M£[/100
pF load,
conductor centered.) |